Model: RODI-2000-03A - High Flow Type I Analytical Grade RO+DI System
This combination Reverse Osmosis System and 8-12 Liters/minute DI System is utilized in operations using virtually any quality tap water. The system is mounted on a welded aluminum frame and includes a built-in three stage pre-filter assembly.
The RO portion of the system will economically remove >97% of the contaminants from the incoming water, and will produce RO pretreated water at the rate of ~ 2-GPM. The RO water is stored in a 200-Liter pressurized storage tank for use by the DI portion of the system.
The pressurized tank eliminates the risk of airborne contamination and the need for a feed pump. The DI portion of the system produces up to 18.2-Megohm-cm water at a flow rate of 8-12 Liters/minute. DI water production is limited by the storage capacity and RO production. This system features continuous recirculation to maintain maximum water purity and a digital conductivity/resistivity meter to monitor the DI water quality. The meter has a green LED to indicate that the water quality is acceptable and a red LED that alerts the user when the water quality drops below the desired level.
All AQUA SOLUTIONS Analytical Grade Type I DI Laboratory Water Purification Systems produce purified water that meets or exceeds ASTM, CAP and CLSI specifications for Type I reagent grade water. Since ordinary tap water contains 5-25 times the impurities of pretreated water, these tap water systems have lower overall capacity than the pretreated water systems.
Features and Benefits
- Made in the USA
- 6,000 - 12,000 Grains as CaCO3 to a 0.02-Megohm end point
- Continuous recirculation is utilized to ensure water quality and prevent the growth of microorganisms within the system
- Dispensing value included
- 0.22-micron (absolute) capsule - Part # - DC2022
- One (1) year warranty
Pure Water Specifications
Resistivity | 18.2 Megohm-cm - Ultrapure |
Conductivity | 0.056 MicroSeimens/cm - Ultrapure |
Ultrapure Grade | Analytical - Meets or exceeds ASTM, CAP, CLSI, and USP Specifications for Type I water |
TOC | < 50-PPB |
Flow Rate | 8-12 Liters/minute |
Particulates | 0.1-Micron) |
Bacteria | - |
Pyrogen/Endotoxin | N/A |
RNase | N/A |
DNase | N/A |
RO Water Specifications
RO Production | 2-Gallons/Minute |
Storage Capacity | RODI2-200: 200-Liter Pressurized Tank - Diameter - 26 in (660 mm) Height - 48 in (1219mm) |
Feed Water
Feed Water Quality | For operation on any quality tap water containing less than 1,000 PPM of total dissolved solids. |
Water Connection | User supplied 3/4" line with 1/2" or 3/4" valve |
Inlet Connection | 3/4" NPTF or NPTM |
Optional Upgrades
UV2006S8 | UV Sterilizer - Prevents the growth of microorganisms within the system |
UV2006X8 | Dual wavelength UV Oxidizer/Sterilizer - Reduces TOC levels to |
UF2000 | Ultrafilter - Removes Pyrogen, Endotoxin, RNase, & DNase |
System Specifications
RO Membranes | Two (2) Part # CR4040-HF4 |
Frame dimensions | Width - 36 in (914 mm), Depth - 19 in (482 mm) Height - 48 in (1219 mm) |
Mounting | Floor |
Operating weight | 280 lbs (127 kg) System/Frame only |
Input Power | 120 or 240 VAC -15% +10%, 50/60 Hz, 25 Watts. |
Minimum Inlet pressure | 35 PSI (2.4 bar) |
Maximum inlet pressure | 100 PSI (6.6 bar) |
Total shipping weight | 500-600 lbs (227 - 272 kg) |
Consumables
CNBB7525MP | 20” 75/25µ prefilter |
CNBB2005PP | 20” 5µ prefilter |
CCBB1030 | 20" 3µ activated carbon |
CR4040-HF4 | High Flow, Low Pressure TFC Membrane |
2635M1T-A | Required Purification Kit - Includes a 2635TO High purity, High capacity DI module and a DC2022 0.22-Micron Final Filter |
2635M1L-A | N/A |
2635M2T-A | Required Purification Kit - Includes two 2635TO High purity, High capacity DI modules and a DC2022 0.22-Micron Final Filter |
2635M2L-A | N/A |
DC2022 | 0.22-Micron Final Filter |
UV2006LS8 | N/A |
UV2006LX8 | N/A |
UF2000RM |