Model: 2035BL - Biological Grade Type I DI System
This AQUA SOLUTIONS unit operates on pretreated water. For ordinary tapwater operations, see Model: 2036B
Laboratory water purification systems are also available with a built-in sterilizer (Model: 2035BU) or with a built-in Dual wavelength UV Oxidizer/Sterilizer (Model: 2035BL).
The Purification Kit Number 2635M1T is required to ensure optimal performance of this laboratory water purification system.
This system produces 18.2 Megohm-cm type I ultrapure, reagent grade water that is free of Pyrogen, Endotoxin, RNase and DNase at a flow rate of 8-12 Liters/minute. The unit is utilized in operations involving pretreated feed water that contains less than 20 PPM of total dissolved solids. It includes a built-in 5,000 MWCO ultrafilter cartridge and a cabinet that can be bench, shelf or wall-mounted at no extra charge.
All AQUA SOLUTIONS Biological Grade Type I DI Laboratory Water Purification Systems produce purified water that meets or exceeds ASTM, CAP and CLSI specifications for Type I reagent grade water.
Features and Benefits
- Ultrafilter removes Pyrogen, Endotoxin, RNase and DNase
- High ion exchange capacity (6,000 Grains) CaCO3 to a 0.02-Megohm end point
- Continuous recirculation is utilized to ensure water quality and prevent the growth of microorganisms within the system
- Ultrapure water available on demand up to 7 ft. (2 meters) from the system via the included hand held dispenser- Part # 700R
- Final filtration of 0.22 µm (absolute) capsule
- Two (2) year system warranty (US & Canada only)This item is on our GSA Contract # GS-07F-5799R
- Made in USA
Pure Water Specifications
Resistivity | 18.2 Megohm-cm - Ultrapure |
Conductivity | 0.056 MicroSeimens/cm - Ultrapure |
Ultrapure Grade Details | Analytical - Meets or exceeds ASTM, CAP, CLSI, and USP Specifications for Type I water |
TOC | < 50-PPB |
Flow Rate Details | 8-12 Liters/minute |
Particulates | 0.1-Micron |
Bacteria | < 1 cfu/ml |
Pyrogen/Endotoxin | < 0.001 EU/ml |
RNase | < 0.01 ng/ml |
DNase | < 4 pg/µL |
Flow Rate | 8-12 LPM |
Feed Water
Feed Water Quality | For operation on water that has been pretreated via reverse osmosis, distillation or deionization by a central or point-of-use pretreatment system, and contains less than 20 PPM of TDS (total dissolved solids). |
Water Connection | User supplied 1/2" NPTF valve (1/2'" Tube X 1/2" NPTM fitting included with system) |
Inlet Connection | 1/2" Tube fitting (20 ft. of tubing supplied with system) |
Optional Upgrades
UV2006S-2 | Prevents the growth of microorganisms within the system |
UV2006X-2 | Dual wavelength UV Oxidizer/Sterilizer - Reduces TOC levels to < 5 PPB and prevents the growth of microorganisms within the system |
System Specifications
Cabinet Dimensions | Width - 20 in (508 mm), Depth - 12 in (305 mm) Height - 20 in (508 mm) |
Mounting | Bench, Shelf, or Wall (no additional charge) |
Operating Weight | 75 - 100 lbs (34 - 45 kg) cabinet only |
Input Power | Requires one 100-240 AC, 50/60Hz, 1.0A grounded receptacle within 5' of the right side of the system cabinet |
System Power Source | Switching power supply - 12 VDC at 2.5 Amps |
Minimum Inlet Pressure | 30 psi (2 bar) |
Maximum Inlet Pressure | 100 PSI (6.6 bar) |
Total shipping weight | 50 lbs (23 kg) |
Consumables
2635M1T | Required Purification Kit - Includes a 2613TO High purity, High capacity DI module, a CN1001MP Sediment Pre-filter, and a DC2022 0.22-Micron Final Filter |
(2) UF200RM-Q | 0.05-Micron (5,000 MWCO) Ultrafilter - Removes Pyrogen, Endotoxin, RNase & DNase |
DC2022 | 0.22 micron (absolute) capsule |
CC1050 | 10" 1-Micron Activated Carbon Prefilter |