Model RO2003 Standard System
A single RO system can be used to pretreat the tap water feed for several Type I or Type II DI lab water purification systems, and/or point-of-use locations.
A Reverse Osmosis System should be considered under the following circumstances:
- Incoming tap water contains more than 170 parts per million (ppm) of total dissolved solids.
- DI water usage exceeds 40 liters per day on a Type II DI System.
- DI water usage exceeds 20 liters per day on a Type I DI System.
There are two criteria to consider when determining if a reverse osmosis pretreatment system is either required, or can be justified, based on cost savings:
- Can the DI system alone produce the desired quality and quantity of purified water that is needed for the applications at hand? IF NOT, AN RO PRETREATMENT SYSTEM IS REQUIRED.
- Would it cost more overall (including capital and operating costs over 1-3 years), to process the tap water via DI alone, or with a combination of RO plus DI?
IF DI ALONE COSTS MORE, AN RO SYSTEM IS JUSTIFIED.
This system is also available for a high flow, high capacity applications: Model RO2000-04.
This compact Laboratory Reverse Osmosis ("RO") System is for operation on virtually any quality tap water, containing up to 1,000 PPM of TDS. It economically produces 30-Liters/hour of RO purified water by removing >95% of the contaminants in the incoming tap water.The system includes a built-in, high-capacity activated carbon pre-filter and a 130-Liter pressurized tank to store the RO purified water
AQUA SOLUTIONS' Reverse Osmosis Pretreatment Systems include a pressurized RO storage tank, which eliminates the need for level controls, overflow piping, and transfer pumps.
Features and Benefits:
- RO production is 30-Liters Per Hour.
- Includes 130-Liter pressurized storage tank reduces the risk of contamination and eliminates the need for a transfer pump.
- Built-in 0.1-Micron Activated Carbon Pre-filter.
- Made in the USA.
- The System has a built in TDS Meter to monitor water quality.
- One (1) year warranty.
- This item is on our GSA Contract # GS-07F-5799R.
Pure Water Specifications
Resistivity 25°C: | >2-Megohm-cm at 25°C |
Conductivity | >0.500 MicroSeimens/cm at 25°C |
Pure Water Grade | Type II |
Flow Rate Details | 3-LPM |
Feed Water
Feed Water Quality | Tap water containing 1,000 PPM of TDS |
Water Connection | User supplied 1/4" or 3/8" NPTF valve (3/8" Tube X 1/4" NPTM & 3/8'" Tube X 3/8" NPTM fittings included with system) |
Inlet Connection | 1/4" Tube fitting (20 ft. of 1/4" of tubing supplied with system) |
Features and Benefits
RO Production | 30-Liters Per Hour |
Pressurized Storage | The included 130-Liter pressurized storage tank reduces the risk of contamination and eliminates the need for a transfer pump. |
Pre-Filter | Built-in 0.1-Micron Activated Carbon |
Country of Origin | Made in the USA |
TDS Monitor Details | The System has a built in TDS Meter to monitor water quality. |
Warranty | 1 Year |
GSA | This item is on our GSA Contract # GS-07F-5799R |
Optional Upgrades
CH1004-PF | 10" Pre-filter assembly with 5-Micron sediment filter |
CH1004PAS | 10" Anti-Scaling Prefilter Assembly |
2700N | Remote Dispenser kit - non-recirculating type - with 0.1 micron final filter capsule |
2700N-NFC | Remote Dispenser kit - non-recirculating type - no final filter capsule |
DC2002 | 1 inch 0.1-Micron (absolute) Final Filter Capsule |
DC2004 | 2 inch 0.1-Micron (absolute) Final Filter Capsule |
DC2022 | High flow 0.22-Micron (absolute) Filter Capsule |
UV2006S | UV Sterilizer - Prevents the growth of microorganisms within the system. |
ROT-200TU-130 | Tank upgrade: Increases storage capacity from 130 to 200 Liters |
ROT-200TU-042 | Tank upgrade: Increases storage capacity from 42 to 200 Liters |
ROT-200 | 200-Liter pressurized storage tank with valve assembly |
System Specifications
RO Membranes | Three (3) Part # CR1812H1 |
Cabinet Dimensions | Width - 20 in (508 mm), Depth - 12 in (305 mm) Height - 20 in (508 mm) |
Mounting | Bench, Shelf, or Wall (no additional charge) |
Operating Weight | 75 - 100 lbs (34 - 45 kg) cabinet only |
Input Power | Requires a 120 VAC electrical outlet within 5’ of the right side of the system. |
Minimum Inlet Pressure | 25 PSI (1.7 bar) |
Maximum Inlet Pressure | 100 PSI (6.6 bar) |
Total Shipping Weight | 100-150 lbs (45 - 68 kg) |
Consumables
CC1050 | 10" 1-Micron Activated Carbon Prefilter |
CN1050MP | 0" 5-Micron Sediment Pre-Filter - For use with optional CH1004-PF |
CR1812H1 | (3) 12" TFC Reverse Osmosis Replacement Cartridge |